Before You Begin
Real-time shelf-life time point testing should be done in separate tests for each time point. Please refer to the
guidelines for details.
The assumption in this workflow is that you will have a marked control available at every time point for panelists to compare the test samples against. If you will not have control available for your testing (i.e. if you will be using acceptance testing method, or intensity scale), you will have to alter the steps to expand the questionnaire to collect information on multiple attributes as per your shelf-life testing requirements. Please
contact Compusense Support for assistance as needed.
The marked control sample should not be included inside any
time point tests. Panelists are not expected to collect data for the
sample that is clearly marked as a control, but rather simply use it to compare other
samples against it. Therefore, it should be presented to panelists as a clearly marked control, but should not be included in the test for data collection on it.
Your time point 0, however, will have to have a blind control with the sample type set to "control". Detailed instructions are outlined in this workflow. The blind control set as a "control" sample type will be used in the analysis to compare all the other time points (samples) against it. Other time point tests will have to have sample types set to something other than "control" for analysis to work properly.